ASHCROFT 251009SW02BXLLVAC

מדריך הוראות למד ואקום של ASHCROFT

Model: 251009SW02BXLLVAC

1. הקדמה

This manual provides essential information for the proper installation, operation, and maintenance of your ASHCROFT Vacuum Gauge. This test gauge is designed for precise vacuum measurement, featuring a durable construction suitable for industrial applications.

The gauge has a 2-1/2 inch dial, measures vacuum from 30 inches of Mercury (Hg) to 0, and features a 1/4 inch NPT connection located at the lower back. It is constructed with a 304 stainless steel case, 316 stainless steel socket and tube, and a polycarbonate lens. The gauge is filled with glycerin for dampening and enhanced performance.

ASHCROFT Vacuum Gauge

איור 1: ASHCROFT Vacuum Gauge. This image displays the front view of the vacuum gauge, highlighting its stainless steel casing, white dial with black markings, and the ASHCROFT DURALIFE branding. The gauge face indicates measurements in psi and notes "Welded AISI 316 tube & socket U.S.A."

2. מידע בטיחות

Always observe standard safety precautions when handling and installing pressure and vacuum gauges. Failure to do so may result in injury or damage to equipment.

  • Ensure the system is depressurized before installing or removing the gauge.
  • Do not exceed the specified operating range of 30" Hg Vac to 0.
  • Verify compatibility of the gauge's wetted parts (316 Stainless Steel) with the process media.
  • Protect the gauge from extreme temperatures outside the -40°F to 150°F operating range.
  • Handle with care to avoid dropping or impacting the gauge, which can affect accuracy.

3. התקנה והתקנה

Proper installation is crucial for accurate readings and long service life.

  1. הֲכָנָה: Ensure the connection port on your system is clean and free of debris.
  2. קֶשֶׁר: The gauge features a 1/4 inch NPT (National Pipe Taper) connection. Apply appropriate thread sealant (e.g., PTFE tape) to the male threads of the gauge connection.
  3. הַרכָּבָה: Carefully thread the gauge into the system's female NPT port. Tighten using a wrench on the wrench flats of the gauge connection. Do not overtighten or apply torque to the gauge case.
  4. הִתמַצְאוּת: The connection location is lower back, allowing for direct mounting or panel mounting with appropriate accessories (not included). Ensure the dial is easily visible for reading.

4. הוראות הפעלה

The ASHCROFT Vacuum Gauge provides accurate vacuum measurements within its specified range.

  • קריאת המד: The dial is marked from 30" Hg Vac (full vacuum) to 0 (atmospheric pressure). The smallest graduation is 0.5" Hg, allowing for precise readings.
  • דִיוּק: This gauge has an accuracy of +/-1.0% of span, ensuring reliable measurements for test applications.
  • שיקולי טמפרטורה: The gauge is designed to operate effectively within a temperature range of -40°F to 150°F. Operating outside this range may affect accuracy and longevity.
  • Glycerin Fill: The glycerin fill dampens pointer pulsation caused by vibration and pressure surges, improving readability and extending the life of the gauge movement.

5. תחזוקה

Regular maintenance ensures the longevity and accuracy of your vacuum gauge.

  • ניקוי: Clean the exterior of the gauge, especially the polycarbonate lens, with a soft, damp cloth. Avoid abrasive cleaners or solvents that could damage the lens or case.
  • בְּדִיקָה: Periodically inspect the gauge for any signs of damage, such as cracks in the lens, dents in the case, or bent pointer. Check the connection for leaks.
  • כיול מחדש: While designed for long-term accuracy, professional recalibration may be necessary if the gauge is subjected to severe conditions or if accuracy is critical for your application.
  • אִחסוּן: When not in use, store the gauge in a clean, dry environment within its operating temperature range.

6. פתרון תקלות

עיין בבעיות הנפוצות הבאות ובפתרונות האפשריים שלהן:

בְּעָיָהסיבה אפשריתפִּתָרוֹן
קריאה לא מדויקתGauge exposed to over-vacuum; physical damage; extreme temperature.Verify operating conditions; inspect for damage; allow gauge to normalize temperature. Consider professional recalibration or replacement.
Pointer FluctuationSystem vibration; rapid pressure changes.The glycerin fill helps dampen this; ensure gauge is securely mounted. If severe, consider a snubber or remote mounting.
Leak at ConnectionImproperly applied thread sealant; insufficient tightening.Depressurize system, remove gauge, reapply thread sealant, and re-install, ensuring proper tightening.

7. מפרטים

Detailed technical specifications for the ASHCROFT Vacuum Gauge (Model: 251009SW02BXLLVAC).

  • סוג מד: מִבְחָן
  • גודל חוגה: 2-1/2 אינץ'
  • לָנוּעַ: 30" Hg Vac to 0
  • גודל חיבור: 1/4" NPT
  • מיקום החיבור: גב תחתון
  • Smallest Graduation: 0.5" Hg
  • דִיוּק: +/-1.0%
  • חומר המארז: 304 נירוסטה
  • צבע מארז: כֶּסֶף
  • חומר שקע: 316 נירוסטה
  • חומר צינור: 316 נירוסטה
  • לְמַלֵא: גליצרין
  • חומר עדשה: פוליקרבונט
  • Ring Material: נירוסטה
  • טווח טמפרטורות עבודה: -40°F עד 150°F
  • סִדרָה: 1009
  • מידות המוצר: 3 x 3 x 3 אינץ'
  • משקל פריט: 1 פאונד
  • ארץ מוצא: אַרצוֹת הַבְּרִית
  • UPC: 094706106517

8. אחריות ותמיכה

This ASHCROFT Vacuum Gauge comes with a אחריות יצרן לשנה. For warranty claims or technical support, please contact ASHCROFT customer service. Retain your proof of purchase for warranty validation.

For further assistance or detailed technical inquiries, please refer to the official ASHCROFT webאתר או צרו קשר עם המפיצים המורשים שלהם.

Ask a question about this manual